Please use this identifier to cite or link to this item:
http://hdl.handle.net/10397/2364
Title: | Evaluation of bone-tendon junction healing using water jet ultrasound indentation method | Authors: | Lu, MH Zheng, YP Lu, HB Huang, QH Qin, L |
Issue Date: | Nov-2009 | Source: | Ultrasound in medicine and biology, Nov. 2009, v. 35, no. 11, p. 1783-1793 | Abstract: | The re-establishment of bone-tendon junction (BTJ) tissues with the junction, characterized as a unique transitional fibrocartilage zone, is involved in many trauma and reconstructive surgeries. Experimental and clinical findings have shown that a direct BTJ repair requires a long period of immobilization, which may be associated with a postoperative weak knee. Therefore, it is necessary to evaluate the morphologic and mechanical properties of BTJ tissues in situ to better understand the healing process for the purpose of reducing the adverse effects of immobilization. We previously reported a noncontact ultrasound water jet indentation system for measuring and mapping tissue mechanical properties. The key idea was to utilize a water jet as an indenter as well as the coupling medium for high-frequency ultrasound. In this article, we used ultrasound water jet indentation to evaluate the BTJ healing process. The system’s capability of measuring the material elastic modulus was first validated using tissue-mimicking phantoms. Then it was employed to assess the healing of the BTJ tissues after partial patellectomy over time on twelve 18-week-old female New Zealand White rabbits. It was found that in comparison with the normal control samples, the elastic modulus of the fibrocartilage of the postoperative samples was significantly smaller, while its thickness increased significantly. Among the postoperative sample groups, the elastic modulus of the fibrocartilage of the samples harvested at week 18 was significantly higher than those harvested at week 6 and week 12, which was even comparable with the value of the control samples at the same sacrifice time. The results suggested that the noncontact ultrasound water jet indentation system provided a nondestructive way to evaluate the material properties of small animal tissues in situ and thus had the ability to evaluate the healing process of BTJ. | Keywords: | Ultrasound Indentation Ultrasound indentation Water jet Bone-to-tendon junction repair Articular cartilage Soft tissue Elasticity |
Publisher: | Elsevier | Journal: | Ultrasound in medicine and biology | ISSN: | 0301-5629 | EISSN: | 1879-291X | DOI: | 10.1016/j.ultrasmedbio.2009.06.1093 | Rights: | Ultrasound in Medicine and Biology © 2009 World Federation for Ultrasound in Medicine & Biology. The journal web site is located at http://www.sciencedirect.com. |
Appears in Collections: | Journal/Magazine Article |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Lu et al UMB 2009 water jet for bone-tendon junction assessment.pdf | Pre-published version | 561.16 kB | Adobe PDF | View/Open |
Page views
113
Last Week
1
1
Last month
Citations as of May 28, 2023
Downloads
185
Citations as of May 28, 2023
SCOPUSTM
Citations
11
Last Week
0
0
Last month
0
0
Citations as of Jun 1, 2023
WEB OF SCIENCETM
Citations
9
Last Week
0
0
Last month
0
0
Citations as of Jun 1, 2023

Google ScholarTM
Check
Altmetric
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.