| Title: | A systematic study of independently-tuned room-specific PBS beam model in a beam-matched multiroom proton therapy system | Authors: | Huang, YH Fang, C Yang, T Cao, L Zhang, G Qu, B Zhang, Y Wang, Z Xu, S |
Issue Date: | 2021 | Source: | Radiation Oncology, 2021, v. 16, no. 1, 206 | Abstract: | Background: In the existing application of beam-matched multiroom proton therapy system, the model based on the commissioning data from the leading treatment room was used as the shared model. The purpose of this study is to investigate the ability of independently-tuned room-specific beam models of beam-matched gantries to reproduce the agreement between gantries’ performance when considering the errors introduced by the modeling process. Methods: Raw measurements of two gantries’ dosimetric characteristics were quantitatively compared to ensure their agreement after initially beam-matched. Two gantries’ beam model parameters, as well as the model-based computed dosimetric characteristics, were analyzed to study the introduced errors and gantries’ post-modeling consistency. We forced two gantries to share the same beam model. The model-sharing patient-specific quality assurance (QA) tasks were retrospectively performed with 36 cancer patients to study the clinical impact of beam model discrepancies. Results: Intra-gantry comparisons demonstrate that the modeling process introduced the errors to a certain extent indeed, which made the model-based reproduced results deviate from the raw measurements. Among them, the deviation introduced to the IDD curves was generally larger than that to the beam spots during modeling. Cross-gantry comparisons show that, from the beam model perspective, the introduced deviations deteriorated the high agreement of the dosimetric characteristics originally shown between two beam-matched gantries, but the cross-gantry discrepancy was still within the clinically acceptable tolerance. In model-sharing patient-specific QA, for the particular gantry, the beam model usage for intensity-modulated proton therapy (IMPT) QA plan generation had no significant effect on the actual delivering performance. All reached a high level of 95.0% passing rate with a 3 mm/3% criterion. Conclusions: It was preliminary recognized that among beam-matched gantries, the independently-tuned room-specific beam model from any gantry is reasonable to be chosen as the shared beam model without affecting the treatment efficacy. |
Keywords: | Beam analysis Beam-matching Monte Carlo Pencil beam scanning Proton therapy |
Publisher: | BioMed Central | Journal: | Radiation oncology | EISSN: | 1748-717X | DOI: | 10.1186/s13014-021-01932-0 | Rights: | © The Author(s) 2021. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data. The following publication Huang, YH., Fang, C., Yang, T. et al. A systematic study of independently-tuned room-specific PBS beam model in a beam-matched multiroom proton therapy system. Radiat Oncol 16, 206 (2021) is available at https://doi.org/10.1186/s13014-021-01932-0 |
| Appears in Collections: | Journal/Magazine Article |
Files in This Item:
| File | Description | Size | Format | |
|---|---|---|---|---|
| Huang_systematic_study_independely.pdf | 1.22 MB | Adobe PDF | View/Open |
Page views
106
Last Week
0
0
Last month
Citations as of Nov 9, 2025
Downloads
36
Citations as of Nov 9, 2025
SCOPUSTM
Citations
2
Citations as of Dec 19, 2025
WEB OF SCIENCETM
Citations
2
Citations as of Dec 18, 2025
Google ScholarTM
Check
Altmetric
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.