Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/1977
PIRA download icon_1.1View/Download Full Text
Title: Objective assessment of limb tissue elasticity : development of a manual indentation procedure
Authors: Zheng, YP 
Mak, AFT
Lue, B
Issue Date: Apr-1999
Source: Journal of rehabilitation research and development, Apr. 1999, v. 36, no. 2, p. 71-85
Abstract: An ultrasound indentation system with a pen-size hand-held probe was developed and used to obtain the effective Young's moduli of forearm and lower limb soft tissues in 12 subjects. Since the probe is manually driven, the alignment of the probe and control of the rate of indentation are parameters upon which the results obtained depend. This paper addresses whether manual indentation tests with the probe are sufficiently acceptable and repeatable for objective biomechanical characterization of limb tissues. Forearms of three normal subjects were tested in two states of muscular contraction. Six different indentation rates, ranging from 0.75 mm/s to 7.5 mm/s, were imposed. The load-indentation responses obtained were shown to be well represented by quadratic functions. A linear elastic indentation solution was used to extract the effective Young's modulus. The material parameters extracted were repeatable and rather rate-insensitive for the range of rates used. The effective Young's modulus obtained was found to significantly increase as a result of contraction of the underlying muscles. Indentor misalignment experiments demonstrated that misalignment affects the measurement from which the effective Young's modulus of soft tissues is calculated. This effect, however, was found to decrease as the tissue thickness increased. With the investigation of the above issues, a procedure has been established for the extraction of effective Young's moduli of limb soft tissues from manual cyclic indentation responses. Tests on experimental subjects' lower limbs further demonstrated that the ultrasonic indentor is a feasible instrument for characterization of the biomechanical properties of limb soft tissues. Paired-t tests showed that the effective Young's moduli of the lower limb soft tissues of three elderly persons with transtibial amputation were significantly smaller than those of six unimpaired young subjects.
Keywords: CAD/CAM
Indentation test
Prosthetics tissue
Biomechanics
Residual limb assessment
Soft tissue biomechanics
Tissue mechanics
Ultrasound instrumentation
Publisher: U.S. Department of Veterans Affairs, Rehabilitation Research and Development Service
Journal: Journal of rehabilitation research and development 
ISSN: 0748-7711
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
1999 J Rehab Res Dev - assessment of limb tissue elasticity.pdf1.02 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Version of Record
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

221
Last Week
0
Last month
Citations as of Apr 21, 2024

Downloads

243
Citations as of Apr 21, 2024

SCOPUSTM   
Citations

123
Last Week
0
Last month
Citations as of Apr 19, 2024

WEB OF SCIENCETM
Citations

103
Last Week
1
Last month
1
Citations as of Apr 25, 2024

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.