Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/108747
PIRA download icon_1.1View/Download Full Text
Title: Scatter radiation distribution to radiographers, nearby patients and caretakers during portable and pediatric radiography examinations
Authors: Tam, SY 
Fung, YY 
Lau, SY 
Lam, WN 
Wong, ETH 
Issue Date: Jul-2023
Source: Bioengineering, July 2023, v. 10, no. 7, 779
Abstract: Scatter radiation from portable and pediatric X-rays could pose a risk to radiographers, nearby patients, and caretakers. We aim to evaluate the spatial scatter radiation distribution to the radiographers, nearby patients, and caretakers during common projections in portable and pediatric X-rays. We evaluated the three-dimensional scatter dose profiles of four and three commonly used portable and pediatric X-ray projections, respectively, by anthropomorphic phantoms and scatter probes. For portable X-ray, the AP abdomen had the highest scatter radiation dose recorded. Radiographer scatter radiation doses were 177 ± 8 nGy (longest cord extension) and 14 ± 0 nGy (hiding behind the portable X-ray machine). Nearby patient scatter radiation doses were 3323 ± 28 nGy (40 cm bed distance), 1785 ± 50 nGy (80 cm bed distance), and 580 ± 42 nGy (160 cm bed distance). The AP chest and abdomen had the highest scatter radiation dose in pediatric X-rays. Caretaker scatter radiation doses were 33 ± 1 nGy (50 cm height) and 659 ± 7 nGy (140 cm height). Although the estimated lens doses were all within safe levels, the use of shielding and caution on dose estimation by inverse square law is suggested to achieve the ALARA principle and dose optimization.
Keywords: Caretaker
Medical radiation dose
Pediatric radiography
Portable radiography
Radiation protection
Radiographer
Scattered radiation
X-ray examination
Publisher: MDPI AG
Journal: Bioengineering 
EISSN: 2306-5354
DOI: 10.3390/bioengineering10070779
Rights: © 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
The following publication Tam S-Y, Fung Y-Y, Lau S-Y, Lam W-N, Wong ET-H. Scatter Radiation Distribution to Radiographers, Nearby Patients and Caretakers during Portable and Pediatric Radiography Examinations. Bioengineering. 2023; 10(7):779 is available at https://doi.org/10.3390/bioengineering10070779.
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
bioengineering-10-00779-v3.pdf6.06 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Version of Record
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

71
Citations as of Nov 10, 2025

Downloads

96
Citations as of Nov 10, 2025

SCOPUSTM   
Citations

7
Citations as of Dec 19, 2025

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.