Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/107087
PIRA download icon_1.1View/Download Full Text
Title: Non-invasive measurement for cardiac variations using a fiber optic sensor
Authors: Lyu, W 
Xu, W
Yang, F
Chen, S 
Tan, F 
Yu, C 
Issue Date: 15-Sep-2021
Source: IEEE photonics technology letters, 15 Sept. 2021, v. 33, no. 18, p. 990-993
Abstract: Cardiovascular diseases (CVDs) are very common in modern society, such as atherosclerosis, hypertension, etc., which have a great impact on heart function. Therefore, hospitalization monitoring alone is far from enough. Long-term monitoring of the heart is needed in daily life. The technology of non-invasive monitoring of the cardiovascular system can meet the needs of long-term monitoring of the heart condition, helping to promote the improvement of lifestyle and daily care, reducing the overall risk of developing CVDs. The purpose of this study is to investigate the cardiac response after different exercises using a 3 × 3 demodulation scheme-based ballistocardiography (BCG) monitoring system. A fiber optic sensor (FOS)-based smart cushion is used to replace the traditional inconvenient electrocardiogram (ECG) for heart rate variability (HRV). The correlation between BCG inter-beat interval (IBI) and ECG IBI is 0.9862, and the RMSE is 0.0139. The BCG signal can assess cardiac contractility by analyzing RJ interval with ECG, which is a practical alternative to the pre-ejection period (PEP).
Keywords: Biomedical transducers
Fiber optic sensors
Optical fiber interference
Optical fibers
Publisher: Institute of Electrical and Electronics Engineers
Journal: IEEE photonics technology letters 
ISSN: 1041-1135
EISSN: 1941-0174
DOI: 10.1109/LPT.2021.3078757
Rights: © 2021 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
The following publication W. Lyu, W. Xu, F. Yang, S. Chen, F. Tan and C. Yu, "Non-Invasive Measurement for Cardiac Variations Using a Fiber Optic Sensor," in IEEE Photonics Technology Letters, vol. 33, no. 18, pp. 990-993, 15 Sept. 2021 is available at https://doi.org/10.1109/LPT.2021.3078757.
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
Lyu_Non-Invasive_Measurement_Cardiac.pdfPre-Published version1.14 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Final Accepted Manuscript
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

4
Citations as of Jun 30, 2024

Downloads

2
Citations as of Jun 30, 2024

SCOPUSTM   
Citations

21
Citations as of Jun 21, 2024

WEB OF SCIENCETM
Citations

15
Citations as of Jun 27, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.