Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/106326
PIRA download icon_1.1View/Download Full Text
Title: Verticalized-tip trajectory tracking of a 3D-printable soft continuum robot : enabling surgical blood suction automation
Authors: Lai, J 
Huang, K 
Lu, B
Zhao, Q 
Chu, HK 
Issue Date: Jun-2022
Source: IEEE/ASME transactions on mechatronics, June 2022, v. 27, no. 3, p. 1545-1556
Abstract: Soft-bodied robotic manipulators have great potential for use in minimally invasive surgery, owing to their advantages of high flexibility with infinite degrees of freedom (DOF). One of the potential applications is to perform blood suctioning, which is inevitable during the surgery. To attain higher efficiency in suctioning, the robotic tip should remain vertical while moving along on the work surface. Motivated by this application, this article presents a novel soft robot design and its control scheme to properly configure the tip of a two-segment soft robot while following a planned trajectory on the work surface. Aiming to reduce the incision size and the possibility of infection, a 3D-printed soft-bodied manipulator utilizing the cable-driven mechanism with a diameter of 9 mm was designed and fabricated. An additional DOF was added through a motorized insertion stage. The robot system was modeled using piecewise constant-curvature assumption, and an RGB-D vision was employed to enhance the accuracy of the kinematic-based controller. Performances of the tip positioning and verticalizing were evaluated via simulation, and further verified through experiments. The results confirm that the manipulator is capable of following different trajectories at various velocities while keeping its tip vertical. Compared to other similar works, our results are satisfactory with an root-mean-square error of trajectory tracking within 7 mm, and a maximum angular deviation of 6°. Fluid suction experiments were conducted to demonstrate its effectiveness for automated 3-D suction. This article offers a new tool to support the surgeons for surgical blood suction.
Keywords: Soft robot
Surgical robot
Trajectory tracking
Publisher: Institute of Electrical and Electronics Engineers
Journal: IEEE/ASME transactions on mechatronics 
ISSN: 1083-4435
EISSN: 1941-014X
DOI: 10.1109/TMECH.2021.3090838
Rights: © 2021 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
The following publication J. Lai, K. Huang, B. Lu, Q. Zhao and H. Chu, "Verticalized-Tip Trajectory Tracking of A 3D-Printable Soft Continuum Robot: Enabling Surgical Blood Suction Automation," in IEEE/ASME Transactions on Mechatronics is available at https://doi.org/10.1109/TMECH.2021.3090838.
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
Lai_Verticalized-Tip_Trajectory_Tracking.pdfPre-Published version17.03 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Final Accepted Manuscript
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

7
Citations as of Jun 30, 2024

Downloads

2
Citations as of Jun 30, 2024

SCOPUSTM   
Citations

20
Citations as of Jul 4, 2024

WEB OF SCIENCETM
Citations

16
Citations as of Jul 4, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.