Source: Ultrasonics, Sept. 2021, v. 116, 106506
Abstract: Matching layer is a critical component that determines the performance of piezoelectric ultrasound transducer. For most piezoelectric materials, their acoustic impedances are significantly higher than human tissues and organs, so a tunable matching layer with a high acoustic impedance is required for optimizing the acoustic wave transmission. In this article, a high compression fabrication method is presented, with which the acoustic impedance of alumina-epoxy composite matching layer can be tuned from 6.50 to 9.47 MRayl by controlling the applied compression pressure and ratio of the components. The maximum acoustic impedance 9.47 MRayl can be achieved by compressing a mixture of 80% alumina weight ratio under a 62.4 MPa pressure. This enhancement mainly relies on the increased acoustic longitudinal velocity which enlarged the tolerance of high to ultra-high frequency transducer fabrication using quarter wavelength matching design. Furthermore, the attenuation of the matching layer developed by this method is only −10 dB/mm at 40 MHz. The very high acoustic impedance value and very low attenuation make this matching material superior than all reported matching materials, and therefore, can enhance the performance of the ultrasound transducers, especially for medical imaging applications.
Keywords: Acoustic impedance enhancement
Alumina-epoxy 0-3 composite
High frequency ultrasound transducer
Matching layer
Publisher: Elsevier
Journal: Ultrasonics 
ISSN: 0041-624X
EISSN: 1874-9968
DOI: 10.1016/j.ultras.2021.106506
Rights: © 2021 Elsevier B.V. All rights reserved.
© 2021. This manuscript version is made available under the CC-BY-NC-ND 4.0 license http://creativecommons.org/licenses/by-nc-nd/4.0/.
The following publication Wong, C.-M., et al. (2021). "Tunable high acoustic impedance alumina epoxy composite matching for high frequency ultrasound transducer." Ultrasonics 116: 106506 is available at https://dx.doi.org/10.1016/j.ultras.2021.106506.
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
Wong_Tunable_High_Acoustic.pdfPre-Published version2.86 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Final Accepted Manuscript
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

59
Last Week
1
Last month
Citations as of Sep 22, 2024

Downloads

122
Citations as of Sep 22, 2024

SCOPUSTM   
Citations

17
Citations as of Sep 26, 2024

WEB OF SCIENCETM
Citations

13
Citations as of Sep 26, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.