Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/94674
PIRA download icon_1.1View/Download Full Text
Title: Three-dimensional intravascular ultrasound imaging using a miniature helical ultrasonic motor
Authors: Wang, B
Zhou, Y
Wang, Y
Li, X
He, L
Wen, Z
Cao, T
Sun, L 
Wu, D
Issue Date: Feb-2022
Source: IEEE transactions on ultrasonics, ferroelectrics, and frequency control, Feb. 2022, v. 60 , no. 2, p. 681-690
Abstract: Existing 3-D intravascular ultrasound (IVUS) systems that combine two electromagnetic (EM) motors to drive catheters are bulky and require considerable efforts to eliminate EM interference (EMI). Here, we propose a new scanning method to realize 3-D IVUS imaging using a helical ultrasonic motor to overcome the aforementioned issues. The ultrasonic motor with compact dimensions (7-mm outer diameter and 30-mm longitudinal length), lightweight (20.5 g), and free of EMI exhibits a great application potential in mobile imaging devices. In particular, it can simultaneously perform rotary and linear motions, facilitating precise 3-D scanning of an imaging catheter. Experimental results show that the signal-to-noise ratio (SNR) of raw images obtained using the ultrasonic motor is 5.3 dB better than that of an EM motor. Moreover, the proposed imaging device exhibits the maximum rotary speed of 12.3 r/s and the positioning accuracy of 2.6 μm at a driving voltage of 240 Vp-p. The 3-D wire phantom imaging and 3-D tube phantom imaging are performed to evaluate the performance of the imaging device. Finally, the in vitro imaging of a porcine coronary artery demonstrates that the layered architecture of the vessel can be precisely identified while significantly increasing the SNR of the raw images.
Keywords: High-frequency transducer
Intravascular ultrasound (IVUS) imaging
Mechanical scanning device
Ultrasonic motor
Publisher: Institute of Electrical and Electronics Engineers
Journal: IEEE transactions on ultrasonics, ferroelectrics, and frequency control 
ISSN: 0885-3010
EISSN: 1525-8955
DOI: 10.1109/TUFFC.2021.3132607
Rights: © 2021 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
The following publication B. Wang et al., "Three-Dimensional Intravascular Ultrasound Imaging Using a Miniature Helical Ultrasonic Motor," in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 69, no. 2, pp. 681-690, Feb. 2022 is available at https://dx.doi.org/10.1109/TUFFC.2021.3132607.
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
Wang_Three-dimensional_Intravascular_Helical.pdfPre-Published version1.2 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Final Accepted Manuscript
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

76
Last Week
0
Last month
Citations as of May 19, 2024

Downloads

171
Citations as of May 19, 2024

SCOPUSTM   
Citations

10
Citations as of May 16, 2024

WEB OF SCIENCETM
Citations

8
Citations as of May 16, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.