Abstract: Study Design: Randomized controlled trial.
Objective: To compare the effectiveness of the automated pressure-adjustable orthosis (PO) and conventional orthosis (CO) for treatment of adolescent idiopathic scoliosis (AIS).
Summary of background data: Orthosis wearing quality may influence its effectiveness for AIS. An automated PO aimed to provide a more optimized and consistent biomechanical environment. Clinical evaluation was conducted to study the effectiveness of this innovative orthosis.
Methods: Patients with AIS who met the criteria (Age 10-14, Cobb 20°-40°, and Risser sign 0-2) were recruited from two clinics and randomly assigned to the PO and CO groups. Compliance sensors were embedded in both groups, while the PO group was set to adjust the interfacial pressure as prescribed automatically. Clinical assessments (radiology and quality of life, QoL) were conducted at the baseline, immediate after orthosis fitting and 1-year follow-up. Orthosis wearing compliance was tracked using thermo and pressure sensors.
Results: Twenty-four patients were enrolled with one drop-out (PO, n = 11; CO, n = 12). Significant immediate in-orthosis correction was observed in the PO (11.0° ± 6.5°, 42.0%, P < 0.001) and CO (10.3° ± 5.3°, 37.6%, P < 0.001) groups. After 1 year, no patient with PO progressed while two with CO had progression more than 5°. The mean daily wearing duration was 1.1 hours longer in the PO group as compared with the CO group (15.4 ± 4.5 vs. 14.3 ± 3.8 h). Moreover, the wearing quality within the targeted pressure was 33.9% higher in the PO group (56.5 ± 16.5% vs. 23.1 ± 12.1%, P < 0.001). No significant difference in the QoL results was observed between two groups nor within both groups during the study period.
Conclusion: This study showed that the automated PO could enhance wearing quality when compared with the CO, thus offering a better biomechanical corrective effect in the study period without adverse effect on the patients' wearing quantity and QoL.
Keywords: adolescent idiopathic scoliosis
Cobb angle
Compliance
Electronic sensor
Interfacial pressure
Mental health
Orthotic treatment
Quality of life
Wearing quality
Wearing quantity
Publisher: Lippincott Williams & Wilkins
Journal: Spine 
ISSN: 0362-2436
EISSN: 1528-1159
DOI: 10.1097/BRS.0000000000003559
Rights: © 2020 Wolters Kluwer Health, Inc. All rights reserved.
This is the accepted version of an article published by Lippincott Williams & Wilkins in Spine. To access the final edited and published work see https://doi.org/10.1097/BRS.0000000000003559
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
Lin_Integlligent-Automated_Pressure-adjustable.pdfPre-Published version2.45 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Final Accepted Manuscript
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

91
Last Week
0
Last month
Citations as of Apr 14, 2025

Downloads

374
Citations as of Apr 14, 2025

SCOPUSTM   
Citations

15
Citations as of Dec 19, 2025

WEB OF SCIENCETM
Citations

12
Citations as of Dec 18, 2025

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.