Title: Comparison of choroidal thickness measurements using semiautomated and manual segmentation methods
Authors: Zhao, M 
Alonso-Caneiro, D
Lee, R 
Cheong, AMY 
Yu, WY 
Wong, HY 
Lam, AKC 
Issue Date: Feb-2020
Source: Optometry and vision science, Feb. 2020, v. 97, no. 2, p. 121-127
Abstract: SIGNIFICANCE This study demonstrated that a semiautomated segmentation method could help inexperienced practitioners to obtain choroidal thickness as good as experienced practitioners.
PURPOSE The purpose of this study was to compare choroidal thickness measurements obtained by semiautomated and manual segmentation methods.
METHODS Optical coherence tomography images of 37 eyes from 37 healthy young subjects acquired by a spectral-domain optical coherence tomography device were reviewed retrospectively. Two naive examiners measured choroidal thickness using manual and semiautomated methods, whereas two experienced examiners used only the semiautomated method. The semiautomated method referred to a fully automated segmentation program customized based on MATLAB and followed manual verification. After highlighting the inner and outer choroidal boundaries through automated segmentation, examiners reviewed these boundaries in each B-scan and conducted manual revisions if segmentation errors occurred. After selecting points where correct boundary was located, the software used a spline fit to blend the corrected region with the rest of the boundary. All measurements were summarized in a 6-mm Early Treatment Diabetic Retinopathy Study grid. Operation time spent to complete retinal and choroidal segmentation on each eye was recorded. Between-examiner agreements, that is, intraclass correlation coefficient and coefficient of reproducibility (CoR), were calculated among four sets of semiautomated measurements, and within-examiner agreements were comparisons between manual and semiautomated results from the same naive examiners. Eyes with thin or thick choroids were also analyzed separately.
RESULTS The between-examiner and within-examiner agreements were excellent with intraclass correlation coefficient of 0.976 or greater. Pairwise within-examiner CoRs ranged from 17.4 to 47.1 μm. Pairwise between-examiner CoRs were between 13.0 and 38.9 μm. Eyes with thin choroid had better agreements than those with thick choroids. On average, naive examiners saved 3 to 5 minutes per eye using the semiautomated method.
CONCLUSIONS With the help of a dedicated software, inexperienced practitioners could obtain choroidal thickness measurements with accuracy similar to experienced practitioners. Processing time with the semiautomated method was also reduced.
Publisher: Lippincott Williams & Wilkins
Journal: Optometry and vision science 
ISSN: 1040-5488
EISSN: 1538-9235
DOI: 10.1097/OPX.0000000000001473
Rights: Copyright © 2020 American Academy of Optometry
This is the accepted version of an article published by Lippincott Williams & Wilkins in Optometry and Vision Science. To access the final edited and published work see https://doi.org/10.1097/OPX.0000000000001473
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
Zhao_Comparison_Choroidal_Thickness.pdfPre-Published version8.32 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Final Accepted Manuscript
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

66
Last Week
0
Last month
Citations as of Nov 10, 2024

Downloads

104
Citations as of Nov 10, 2024

SCOPUSTM   
Citations

7
Citations as of Nov 7, 2024

WEB OF SCIENCETM
Citations

7
Citations as of Nov 7, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.