Title: Application of fibre bragg grating sensors in strain monitoring and fracture recovery of human femur bone
Authors: Najafzadeh, A
Gunawardena, DS
Liu, Z
Tran, T
Tam, HY
Fu, J
Chen, BK
Issue Date: 2020
Source: Bioengineering, 2020, v. 7, no. 3, 98, p. 1-14
Abstract: Fibre Bragg Grating (FBG) sensors are gaining popularity in biomedical engineering. However, specific standards for in vivo testing for their use are absolutely limited. In this study, in vitro experimental tests were performed to investigate the behaviors and applications of gratings attached to intact and fractured thighbone for a range of compression loading (<300 N) based around some usual daily activities. The wavelength shifts and the corresponding strain sensitivities of the FBG sensors were measured to determine their effectiveness in monitoring the femoral fracture healing process. Four different arrangements of FBG sensors were selected to measure strains at different critical locations on the femoral sawbones surface. Data obtained for intact and plated sawbones were compared using both embedded longitudinal and coiled FBG arrays. Strains were measured close to the fracture, posterior linea aspera and popliteal surface areas, as well as at the proximal and distal ends of the synthetic femur; their responses are discussed herein. The gratings on the longitudinally secured FBG arrays were found to provide high levels of sensitivity and precise measurements, even for relatively small loads (<100 N). Nevertheless, embedding angled FBG sensors is essential to measure the strain generated by applied torque on the femur bone. The maximum recorded strain of the plated femur was 503.97 µε for longitudinal and −274.97 µε for coiled FBG arrays, respectively. These project results are important to configure effective arrangements and orientations of FBG sensors with respect to fracture position and fixation implant for future in vivo experiments.
Keywords: Biomechanics
Fibre bragg grating (FBG) strain sensor
Fracture recovery
In vitro bone testing
Thighbone (Femur)
Publisher: MDPI AG
Journal: Bioengineering 
EISSN: 2306-5354
DOI: 10.3390/bioengineering7030098
Rights: © 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
The following publication Najafzadeh A, Serandi Gunawardena D, Liu Z, Tran T, Tam H-Y, Fu J, K. Chen B. Application of Fibre Bragg Grating Sensors in Strain Monitoring and Fracture Recovery of Human Femur Bone. Bioengineering. 2020; 7(3):98, is available at https://doi.org/10.3390/bioengineering7030098
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
Najafzadeh_Application_fibre_bragg.pdf1.54 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Version of Record
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

67
Last Week
0
Last month
Citations as of Sep 22, 2024

Downloads

25
Citations as of Sep 22, 2024

SCOPUSTM   
Citations

11
Citations as of Sep 26, 2024

WEB OF SCIENCETM
Citations

11
Citations as of Jun 20, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.