Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/87945
PIRA download icon_1.1View/Download Full Text
Title: Finite element analysis of subtalar joint arthroereisis on adult-acquired flexible flatfoot deformity using customised sinus tarsi implant
Authors: Wong, DWC 
Wang, Y 
Niu, W
Zhang, M 
Issue Date: 2020
Source: Journal of orthopaedic translation, 2020
Abstract: Background: Subtalar arthroereisis may cause sinus tarsi pain complications. In this study, we aimed to introduce a customised implant that facilitated treatment effect and less impingement. The biomechanical outcome between the intact and implant conditions was compared using finite element analysis.
Methods: A female patient with flatfoot (age: 36 years, height: 156 ​cm, body mass: 51 ​kg) was recruited as the model patient. The customised implant was designed from the extracted geometry. Boundary and loading conditions were assumed from the data of a normal participant. Four gait instants, including the ground reaction force first peak (25% stance), valley (45%), initial push-off (60%) and second peak (75%) were analyzed.
Results: The navicular height was elevated by 4.2% at 25% stance, whereas the strain of the spring, plantar cuneonavicular and plantar cuboideonavicular ligaments were reduced. The talonavicular joint force decreased and the calcaneocuboid joint increased by half and 67%, respectively, representing a lateralised load pathway. There was a stress concentration at the sulcus tali reaching 15.29 ​MPa
Conclusion: Subtalar arthroereisis using a customised implant may produce some positive treatment effects in terms of navicular height elevation, ligament strain relief and lateralised joint loading pathway. Although the concentrated stress at the sulcus tali did not exceed the threshold of bone breakdown, we could not rule out the potential of vascular disturbance owing to the remarkable elevation of stress. Future study may enlarge the contact area of the bone–implant interface by considering customisation based on the dynamic change of the sinus tarsi during walking gait. The translational potential of this article: Geometry mismatch of prefabricated implants could be the reason for complications. With the advancement of 3D printing, customising implant becomes possible and may improve treatment outcome. This study implemented a theoretical model approach to explore its potential under a simulation of walking.
Keywords: Extra-osseous talotarsal stabilisation
Pes planus
Posterior tibial tendon dysfunction
Sinus tarsi implant
Talotarsal mechanism
Publisher: Elsevier
Journal: Journal of orthopaedic translation 
ISSN: 2214-031X
DOI: 10.1016/j.jot.2020.02.004
Rights: © 2020 The Author(s). Published by Elsevier (Singapore) Pte Ltd on behalf of Chinese Speaking Orthopaedic Society. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
The following publication Wong DW-C et al., Finite element analysis of subtalar joint arthroereisis on adult-acquiredflexibleflatfoot deformity usingcustomised sinus tarsi implant, Journal of Orthopaedic Translation, is available at https://doi.org/10.1016/j.jot.2020.02.004
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
1-s2.0-S2214031X20300139-main.pdf2.26 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Version of Record
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

72
Last Week
0
Last month
Citations as of Apr 28, 2024

Downloads

22
Citations as of Apr 28, 2024

SCOPUSTM   
Citations

11
Citations as of Apr 26, 2024

WEB OF SCIENCETM
Citations

10
Citations as of Apr 25, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.