Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/70387
PIRA download icon_1.1View/Download Full Text
Title: Finite element analysis of locking plate and two types of intramedullary nails for treating mid-shaft clavicle fractures
Authors: Ni, M 
Niu, W
Wong, DWC 
Zeng, W
Mei, J
Zhang, M 
Issue Date: Aug-2016
Source: Injury, Aug. 2016, v. 47, no. 8, p. 1618-1623
Abstract: Background: Both plate and intramedullary nail fixations, including straight and anatomic nails, have been clinically adopted for the treatment of displaced mid-shaft clavicle fractures. However, the biomechanical performances of these fixations and implants have not been well evaluated. This study aims to compare the construct stability, stress distribution and fracture micro-motion of three fixations based on finite element (FE) method. Methods: The FE model of clavicle was reconstructed from CT images of a male volunteer. A mid-shaft fracture gap was created in the intact clavicle. Three fixation styles were simulated including locking plate (LP), anatomic intramedullary nail (CRx), and straight intramedullary nail (RCP). Two loading scenarios (axial compression and inferior bending) were applied at the distal end of the clavicle to simulate arm abduction, while the sternal end was fixed. Results: Under both conditions, the LP was the stiffest, followed by the CRx, and the RCP was the weakest. LP also displayed a more evenly stress distribution for both implant and bone. RCP had a higher stress compared with CRx in both conditions. Moreover, all implants sustained higher stress level under the loading condition of bending than compression. Conclusions: The plate fixation significantly stabilizes the fracture gap, reduces the implant stress, and serves as the recommended fixation for the mid-shaft clavicle fracture. The CRx is an alternative device to treat clavicle shaft fracture, but the shoulder excessive activities should be avoided after operation.
Keywords: Biomechanics
Clavicle fracture
Internal fixation
Plate
Intramedullary nailing
Finite element
Publisher: Elsevier
Journal: Injury 
ISSN: 0020-1383
EISSN: 1879-0267
DOI: 10.1016/j.injury.2016.06.004
Rights: © 2016 Elsevier Ltd. All rights reserved.
© 2016. This manuscript version is made available under the CC-BY-NC-ND 4.0 license https://creativecommons.org/licenses/by-nc-nd/4.0/
The following publication Ni, M., Niu, W., Wong, D. W. C., Zeng, W., Mei, J., & Zhang, M. (2016). Finite element analysis of locking plate and two types of intramedullary nails for treating mid-shaft clavicle fractures. Injury, 47(8), 1618-1623 is available at https://doi.org/10.1016/j.injury.2016.06.004
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
Ni_Finite_Element_Analysis.pdfPre-Published version741.97 kBAdobe PDFView/Open
Open Access Information
Status open access
File Version Final Accepted Manuscript
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

86
Last Week
1
Last month
Citations as of Apr 14, 2024

Downloads

99
Citations as of Apr 14, 2024

SCOPUSTM   
Citations

31
Last Week
0
Last month
Citations as of Apr 5, 2024

WEB OF SCIENCETM
Citations

22
Last Week
0
Last month
Citations as of Apr 18, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.