Title: | Comparison between reflection-mode photoplethysmography and arterial diameter change detected by ultrasound at the region of radial artery | Authors: | Wang, C Zheng, YP |
Issue Date: | Aug-2010 | Source: | Blood pressure monitoring, Aug. 2010, v. 15, no. 4, p. 213-219 | Abstract: | Objective: Continuous monitoring of arterial blood pressure (ABP) can provide vital information on the cardiovascular system in clinical practice and research. To achieve this, methods have been developed to determine ABP continuously and noninvasively, including the wrist volume clamp method, which uses reflection-mode photoplethysmography (PPG) to indicate the volume change of the radial artery. However, this kind of indication is reportedly not accurate enough, and an in-vivo test for its accuracy is still lacking. Our aim was to fix this gap. Methods: As ultrasound can detect the arterial diameter accurately; we have developed a sensor comprised a PPG sensor and an ultrasound transducer, which can collect reflection-mode PPG and A-mode ultrasound signals, simultaneously. Tests on 15 volunteers were conducted under the similar conditions of using wrist volume clamp method for a comparison between the PPG signal and the radial arterial diameter. Results: It was observed that there were obvious differences between the shapes of these two signals, mainly between their drifting trend parts, with the pulsatile components of the signals matching very well. The mean root mean square value of the differences of the trend parts (0.287±0.072) was found to be much larger than that of the pulsatile parts (0.107±0.028). Conclusion: The results indicated that using PPG signal as a reference for ABP measurement at the wrist region might not be accurate enough, and ultrasound has the potential for replacing it in the wrist volume clamp method for more accurate ABP determination. |
Keywords: | Arterial diameter Bood pressure Potoplethysmography Reflection-mode photoplethysmography Ultrasound Wrist volume clamp method |
Publisher: | Lippincott Williams & Wilkins | Journal: | Blood pressure monitoring | ISSN: | 1359-5237 | EISSN: | 1473-5725 | DOI: | 10.1097/MBP.0b013e328338aada | Rights: | © 2010 Wolters Kluwer Health|Lippincott Williams & Wilkins. This is a non-final version of an article published in final form in Wang, C., & Zheng, Y. (2010). Comparison between reflection-mode photoplethysmography and arterial diameter change detected by ultrasound at the region of radial artery. Blood Pressure Monitoring, 15(4), 213-219. |
Appears in Collections: | Journal/Magazine Article |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
BPMJ-2009-50 (Pre-PublishedVersion).pdf | Pre-published version | 800.04 kB | Adobe PDF | View/Open |
Page views
176
Last Week
0
0
Last month
Citations as of Apr 28, 2024
Downloads
714
Citations as of Apr 28, 2024
SCOPUSTM
Citations
9
Last Week
0
0
Last month
0
0
Citations as of Apr 4, 2024
WEB OF SCIENCETM
Citations
4
Last Week
0
0
Last month
0
0
Citations as of May 2, 2024
Google ScholarTM
Check
Altmetric
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.