Title: A novel instrument for logging nearwork distance
Authors: Leung, TW
Flitcroft, DI
Wallman, J
Lee, TH
Zheng, Y 
Lam, CSY 
Kee, CS 
Keywords: Instrument
Myopia
Near working distance
Reading habits
Refractive error
Issue Date: 2011
Publisher: Wiley-Blackwell
Source: Ophthalmic and physiological optics, 2011, v. 31, no. 2, p. 137-144 How to cite?
Journal: Ophthalmic and physiological optics 
Abstract: Purpose: To validate a novel ultrasonic sensor for logging reading distances. In addition, this device was used to compare the habitual reading distances between low and high myopes. Methods: First, the stability and sensitivity of the ultrasonic device were determined by repeated measures using artificial targets. Then, thirty Hong Kong Chinese (20-30years) were recruited, of whom fifteen were considered to be high myopes (mean±S.D.=-8.7±0.5D) and 15 to be low to non-myopes (mean±S.D.=-2.0±0.2D). Each subject read a newspaper with their habitual visual aid continuously for 10min in two sessions at their preferred working distance(s). The reading distances were recorded continuously using a novel nearwork analyzer. The modal working distance was considered as the 'habitual' reading distance. In addition, habitual reading distance was reported orally by each subject. Results: The nearwork analyzer gave accurate and repeatable measurements over a range of distances and angles. Using this instrument, high myopes were found to have a significantly shorter reading distance than low myopes or non-myopes (mean±S.D.=35.9±9.8cm vs 50.9±24.8cm; two-sample t-test, p=0.04, df=18). The reading distances reported orally by the subjects were not correlated with those recorded by the nearwork analyzer. Conclusions: The nearwork analyzer was found to be an effective tool for measuring nearwork reading distance in a small group of emmetropic and myopic adults over a 10min interval. Differences between the reading distance between high myopes and low/non-myopes was detected by the device. Further study is needed to determine if a closer working distance is a cause or effect of myopia development.
URI: http://hdl.handle.net/10397/30870
ISSN: 0275-5408
EISSN: 1475-1313
DOI: 10.1111/j.1475-1313.2010.00814.x
Appears in Collections:Journal/Magazine Article

Access
View full-text via PolyU eLinks SFX Query
Show full item record

SCOPUSTM   
Citations

5
Last Week
0
Last month
0
Citations as of Sep 8, 2017

WEB OF SCIENCETM
Citations

5
Last Week
0
Last month
0
Citations as of Sep 22, 2017

Page view(s)

54
Last Week
2
Last month
Checked on Sep 17, 2017

Google ScholarTM

Check

Altmetric



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.