Journal: Journal of rehabilitation research and development 
Abstract: We introduce a method, known as one-dimensional sonomyography (1-D SMG), that uses A-mode ultrasound signals to detect dynamic thickness changes in skeletal muscle during contraction. We custom-designed a 1-D SMG system to collect synchronized A-mode ultrasound, joint angle, and surface electromyography (EMG) signals of forearm muscles during wrist extension. We extracted the 1-D SMG signal from the ultrasound signal by automatically tracking the corresponding echoes, which we then used to calculate muscle thickness changes. We tested the right forearm muscles of nine nondisabled young subjects while they performed wrist extensions at 15.0, 22.5, and 30.0 cycles/min and their largest wrist extension angle ranged from 80° to 90°. We found that the muscle deformation and EMG root mean square signals correlated linearly with wrist extension angle. The ratio of deformation to wrist angle was significantly different among the subjects (p <0.001) but not among the trials of different extension rates for each subject (p = 0.9). The results demonstrate that 1-D SMG can be reliably performed and that it has the potential for skeletal muscle assessment and prosthesis control.
URI: http://hdl.handle.net/10397/1793
ISSN: 0748-7711
DOI: 10.1682/JRRD.2007.02.0026
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
2008 JRRD Dynamic monitoring of forearm muscles usingone-dimensional sonomyography system.pdf387.15 kBAdobe PDFView/Open
Access
View full-text via PolyU eLinks SFX Query
Show full item record

SCOPUSTM   
Citations

20
Last Week
0
Last month
0
Citations as of Apr 23, 2016

WEB OF SCIENCETM
Citations

16
Citations as of Apr 27, 2016

Page view(s)

553
Last Week
2
Last month
Checked on Apr 24, 2016

Download(s)

507
Checked on Apr 24, 2016

Google ScholarTM

Check

Altmetric



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.