| Title: | Laboratory validation of a fully automated point-of-care device for high-order multiplexing real-time PCR detection of respiratory pathogens | Authors: | Li, LCW Tai, DMS Yee, A Tsui, NBY Tsang, PYL Chu, SLH Leung, CT Leung, BKW Wong, W Kebede, FT Leung, PYM Chung, T Yip, CCY Chen, JHK Poon, RWS To, KKW Yuen, KY Fok, M Lau, JYN Lau, LT |
Issue Date: | Oct-2025 | Source: | Diagnostics, Oct. 2025, v. 15, no. 19, 2445 | Abstract: | Background/Objectives: We have previously reported the engineering of a point-of-care (POC) system that fully automates the procedures for nucleic acid extraction and multiplexed real-time RT-PCR, with a major advantage of high-level multiplexing. In this study, we applied and validated the system in a respiratory tract infection setting. Methods: An automatic nested real-time RT-PCR assay was developed (POCm). It was a 40-plex assay that simultaneously detected 39 epidemiologically important respiratory pathogens in 1.5 h in the POC system. The analytical and clinical performance was evaluated. Results: The analytical sensitivities of the POCm assay were comparable to those of its single-plex counterparts performed manually on a bench-top. The minimum detectable concentrations ranged from 53 copies/mL to 5.3 × 103 copies/mL for all pathogen targets except hCoV-NL63 (5.3 × 104 copies/mL). The quantitative performance was demonstrated by the linear correlations between Ct values and input concentrations for all pathogen targets, with 24 of them demonstrating coefficients of correlation (r) greater than 0.9. The POCm assay was subsequently evaluated in 283 clinical samples. A high level of agreement (98.2–100%) was achieved for pathogen detection results between POCm and standard diagnostic methods. The POCm result was also fully concordant with the result of another commercial POC multiplex platform. For positive clinical samples, pairwise Ct values measured by POCm closely correlated with those of the bench-top reference method (r = 0.70). The feasibility of mutation genotyping of the viral subtype was further demonstrated. Conclusions: This study demonstrated the practicality of POCm for routine testing in clinical laboratories. Further clinical trials are being conducted to evaluate the clinical performance of the system. |
Keywords: | Automated genetic pathogen test Multiplex real-time RT-PCR Pathogen detection Point-of-care system Respiratory tract infection |
Publisher: | MDPI AG | Journal: | Diagnostics | EISSN: | 2075-4418 | DOI: | 10.3390/diagnostics15192445 | Rights: | Copyright: © 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). The following publication Li, L. C. W., Tai, D. M. S., Yee, A., Tsui, N. B. Y., Tsang, P. Y. L., Chu, S. L. H., Leung, C. T., Leung, B. K. W., Wong, W., Kebede, F. T., Leung, P. Y. M., Chung, T., Yip, C. C. Y., Chen, J. H. K., Poon, R. W. S., To, K. K. W., Yuen, K.-Y., Fok, M., Lau, J. Y. N., & Lau, L. T. (2025). Laboratory Validation of a Fully Automated Point-of-Care Device for High-Order Multiplexing Real-Time PCR Detection of Respiratory Pathogens. Diagnostics, 15(19), 2445 is available at https://doi.org/10.3390/diagnostics15192445. |
| Appears in Collections: | Journal/Magazine Article |
Files in This Item:
| File | Description | Size | Format | |
|---|---|---|---|---|
| diagnostics-15-02445.pdf | 1.17 MB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.