Title: Investigation of 3D vessel reconstruction under Doppler imaging with phantoms : towards reconstruction of the Circle of Willis
Authors: Li, S 
Shea, QTK 
Ling, YT 
Zheng, YP 
Issue Date: Jul-2024
Source: Ultrasonics, July 2024, v. 141, 107332
Abstract: Background: Stroke is the second leading cause of death across the globe. Early screening and risk detection could provide early intervention and possibly prevent its incidence. Imaging modalities, including 1D-Transcranial Doppler Ultrasound (1D-TCD) or Transcranial Color-code sonography (TCCS), could only provide low spatial resolution or 2D image information, respectively. Notably, 3D imaging modalities including CT have high radiation exposure, whereas MRI is expensive and cannot be adopted in patients with implanted devices. This study proposes an alternative imaging solution for reconstructing 3D Doppler ultrasound geared towards providing a screening tool for the 3D vessel structure of the brain.
Methods: The system comprises an ultrasound phased array attached to a servo motor, which can rotate 180˚ at a speed of 2˚/s. We extracted the color Doppler ROI from the image before reconstructing it into a 3D view using a customized pixel-based algorithm. Different vascular diameters, flow velocity, and depth were tested using a vascular phantom with a pumped flow to confirm the system for imaging blood flow. These variables were set to mimic the vessel diameter, flow speed, and depth of the Circle of Willis (CoW) during a transcranial screening.
Results and conclusions: The lower values of absolute error and ratio were found in the larger vascular channels, and vessel diameter overrepresentation was observed. Under different flow velocities, such diameter overrepresentation in the reconstructed flow did not change much; however, it did change with different depths. Meanwhile, the setting of the velocity scale and the color gain affected the dimension of reconstructed objectives. Moreover, we presented a 3D image of CoW from a subject to demonstrate its potential. The findings of this work can provide a good reference for further studies on the reconstruction of the CoW or other blood vessels using Doppler imaging.
Keywords: 3D ultrasound
Brain
Circle of Willis
Doppler imaging
Stroke
Ultrasound imaging
Publisher: Elsevier BV
Journal: Ultrasonics 
ISSN: 0041-624X
EISSN: 1874-9968
DOI: 10.1016/j.ultras.2024.107332
Rights: © 2024 The Authors. Published by Elsevier B.V. This is an open access article under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
The following publication Li, S., Shea, Q. T. K., Ling, Y. T., & Zheng, Y.-P. (2024). Investigation of 3D vessel reconstruction under Doppler imaging with phantoms: Towards reconstruction of the Circle of Willis. Ultrasonics, 141, 107332 is available at https://doi.org/10.1016/j.ultras.2024.107332.
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
1-s2.0-S0041624X24000945-main.pdf3.27 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Version of Record
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

47
Citations as of Apr 14, 2025

Downloads

27
Citations as of Apr 14, 2025

SCOPUSTM   
Citations

1
Citations as of Sep 12, 2025

WEB OF SCIENCETM
Citations

1
Citations as of Nov 14, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.