Title: Dynamic pressure measurements during vitrectomy in a model of the eye
Authors: Nepita, I
Stocchino, A 
Dodero, A
Castellano, M
Ferrara, M
Romano, MR
Repetto, R
Issue Date: May-2022
Source: Translational Vision Science & technology, May 2022, v. 11, no. 5, 21
Abstract: Purpose: To accurately evaluate pressure changes during vitrectomy in a rigid model of the vitreous chamber and to test the efficiency of the EVA phacovitrectomy system (Dutch Ophthalmic Research Center) in terms of compensation of intraocular pressure variations.
Methods: We tested 23-, 25-, and 27-gauge double-blade vitreous cutters in both vented global pressure control and automatic infusion compensation (AIC) modes in a vitreous chamber model, mimicking the real surgical procedure. Balanced salt solution and artificial vitreous, similar to the real vitreous body, were used. We tested both standard-flow (SF) and high-flow (HF) infusion systems, varying the infusion pressure between 20 and 40 mm Hg. In each experiment, flow rate was also measured.
Results: Pressure drop was rapidly and efficiently compensated when 23-and 25-gauge cutters were used in AIC mode, with infusion pressures ranging between 30 and 55 mm Hg. The 27-gauge cutter was less efficient in compensating pressure variations. Pressure fluctuations related to the high-frequency motion of the cutter blade were small compared to the overall pressure variations. The use of the HF infusion system resulted in larger flow rates and lower pressure changes compared to the SF infusion system.
Conclusions: Despite the rigid material of the model, the present pressure measurements are in line with previous studies performed on porcine eye. The use of AIC mode compensates intraoperative pressure drops efficiently, with both 23-and 25-gauge cutters. The HF infusion system is more efficient than the SF infusion system.
Translational Relevance: The AIC infusion mode efficiently compensates intraopera-tive pressure drops, in both 23-and 25-gauge experimental vitrectomy. The HF infusion system resulted in larger flow rate and lower pressure changes.
Keywords: IOP measurements
Pressure compensation, artificial vitreous
Vitrectomy
Publisher: Association for Research in Vision and Ophthalmology
Journal: Translational vision science & technology 
EISSN: 2164-2591
DOI: 10.1167/tvst.11.5.21
Rights: Copyright 2022 The Authors
This work is licensed under a Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0/).
The following publication Nepita, I., Stocchino, A., Dodero, A., Castellano, M., Ferrara, M., Romano, M. R., & Repetto, R. (2022). Dynamic pressure measurements during vitrectomy in a model of the eye. Translational Vision Science & Technology, 11(5), 21-21 is available at https://doi.org/10.1167/tvst.11.5.21.
Appears in Collections:Journal/Magazine Article

Files in This Item:
File Description SizeFormat 
i2164-2591-11-5-21_1652772105.80949.pdf2.09 MBAdobe PDFView/Open
Open Access Information
Status open access
File Version Version of Record
Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page views

111
Last Week
0
Last month
Citations as of Nov 9, 2025

Downloads

77
Citations as of Nov 9, 2025

SCOPUSTM   
Citations

3
Citations as of Dec 19, 2025

WEB OF SCIENCETM
Citations

3
Citations as of Dec 18, 2025

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.