Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/55963
Title: Is linac-based volumetric modulated arc therapy better than helical tomotherapy in the radiotherapy of nasopharyngeal carcinoma?
Authors: Wu, V 
Choi, BK
Chan, HN
Lam, R
Wong, YK
Mui, AWL
Chiu, G
Keywords: Nasopharyngeal carcinoma
Radiotherapy
Volumetric modulated arc therapy
Helical tomotherapy
Dosimetric comparison
Issue Date: 2015
Publisher: Springer
Source: Journal of radiation oncology, 2015, v. 4, no. 1, p. 29-35 How to cite?
Journal: Journal of radiation oncology 
Abstract: Objective
Volumetric arc therapy (VMAT) and helical tomotherapy (HT) are currently the two state-of-the-art rotational intensity-modulated radiotherapy (IMRT). When compared with conventional IMRT, linac-based VMAT can shorten the treatment time while HT has been claimed to produce a better dosimetric outcome. The aim of our study was to evaluate whether the more “efficient” VMAT can produce a comparable or better dosimetric outcome when compared with HT for the treatment of nasopharyngeal carcinoma (NPC) patients.
Methods
Sixteen primary NPC patients treated by HT were recruited. In addition to the original HT plans, two-arc VMAT plans using the same patient data and dose requirements were generated for each patient. The dosimetric evaluation of the plans included the comparison of conformity index (CI), homogeneity index (HI), dose to 2 and 98 % volume (D2 and D98) of the planning target volume (PTV) of the NP, and left and right lymphatic regions. For the organs at risk (OARs), maximum dose or D2 and mean dose were used for serial and parallel organs, respectively. Student’s paired t test or Wilcoxon signed-rank test was used to check for significant difference.
Results
The dose to the NP target was similar between the two techniques. For the nodal targets, the VMAT plans produced higher mean D2 and lower HI values. For the OARs, the spinal cord and spinal cord doses were higher in the VMAT plans than in the HT plans, whereas the opposite happened in the optic chiasm and pituitary gland. The doses to the brain stem, lens, and optic nerve did not show a significant difference.
Conclusion
Both systems delivered satisfactory dose distributions and kept the dose to OARs below their tolerance. However, if the parotid gland toxicity is a concern in the treatment, HT would be more favorable because it is relatively more effective in sparing the parotid gland. Otherwise, the linac-based double-arc VMAT would be recommended because of its shorter treatment time.
URI: http://hdl.handle.net/10397/55963
ISSN: 1948-7894 (print)
1948-7908 (online)
DOI: 10.1007/s13566-014-0173-4
Appears in Collections:Journal/Magazine Article

Access
View full-text via PolyU eLinks SFX Query
Show full item record

Page view(s)

78
Last Week
1
Last month
Checked on Nov 20, 2017

Google ScholarTM

Check

Altmetric



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.