Title: Anterior chamber angle measurement with Anterior Eye Segment analysis system Nidek EAS-1000® : improving the repeatability
Authors: Lee, TT
Lam, AKC 
Chan, BLK
Keywords: Anterior chamber angle
EAS-1000® anterior eye segment analysis system
Optical axis
Visual axis
Issue Date: 2003
Publisher: Wiley-Blackwell
Source: Ophthalmic and physiological optics, 2003, v. 23, no. 5, p. 423-428 How to cite?
Journal: Ophthalmic and physiological optics 
Abstract: Previous studies have reported good intraobserver and interobserver repeatability of anterior chamber angle (ACA) assessment using the Anterior Eye Segment analysis system, EAS-1000®. This study investigated the influence of fixation pattern, corneal plotting method and camera position on ACA measurements. A Nidek EAS-1000® system was employed for image acquisition and analysis, to obtain measurements of ACA. Twenty-five young healthy subjects participated. The ACAs were different in different quadrants and using different corneal plotting methods (Repeated measures 3-way ANOVA, p < 0.05). However, the difference was <5° and therefore not clinically significant. Alignment along the visual axis or optical axis did not yield any significant difference. Repeatability was improved by asking the subjects to fixate at the flashing light-emitting diode (LED). Temporal and nasal angles also demonstrated better repeatability than vertical angles. In measuring the vertical angles, the camera can be set at either the temporal or nasal side (Paired t-test, p > 0.05). The Anterior Eye Segment analysis system has good clinical repeatability in measuring ACA. It can be further improved with the use of the flashing LED during image acquisition.
URI: http://hdl.handle.net/10397/29100
ISSN: 0275-5408
EISSN: 1475-1313
DOI: 10.1046/j.1475-1313.2003.00133.x
Appears in Collections:Journal/Magazine Article

Access
View full-text via PolyU eLinks SFX Query
Show full item record

SCOPUSTM   
Citations

5
Last Week
0
Last month
0
Citations as of Sep 19, 2017

WEB OF SCIENCETM
Citations

6
Last Week
0
Last month
0
Citations as of Sep 14, 2017

Page view(s)

47
Last Week
4
Last month
Checked on Sep 18, 2017

Google ScholarTM

Check

Altmetric



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.