Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/22091
Title: Improvement of corrosion resistance of pure magnesium in Hanks' solution by microarc oxidation with sol-gel TiO2 sealing
Authors: Shi, P
Ng, WF
Wong, MH
Cheng, FT
Keywords: Corrosion
Magnesium
Microarc oxidation
Simulated body fluid
Sol-gel coating
Issue Date: 2009
Publisher: Elsevier
Source: Journal of alloys and compounds, 2009, v. 469, no. 1-2, p. 286-292 How to cite?
Journal: Journal of alloys and compounds 
Abstract: A composite coating was prepared on degradable magnesium (Mg) implant for improving its corrosion resistance in Hanks' solution (a simulated body fluid). The composite coating was fabricated using a two-step process: (i) a thick and porous oxide layer (with pore size of a few μm) was first formed by microarc oxidation (MAO) in a silicate/fluoride containing electrolyte, and (ii) a top TiO2 sealing layer was formed on the porous layer by sol-gel dip coating followed by hydrothermal treatment. SEM imaging and EDS mapping revealed an average overall thickness of about 12 μm. XRD analysis showed that the MAO layer was composed of magnesium oxide and fluoride and the TiO2 was present as an amorphous phase. Both electrochemical impedance spectroscopic (EIS) and anodic polarization measurements consistently recorded an increase in the initial corrosion resistance of about 30 times due to the composite coating. Immersion tests also showed that the corrosion behavior of the coated samples was more stable over time, and the degree of corroded damage was much reduced compared with bare Mg. The porous MAO layer provided anchorage sites for the subsequent TiO2 sol-gel coating, which sealed the pores and hence significantly enhanced the corrosion resistance while direct sol-gel coating on bare Mg was not successful. The present results indicate that the corrosion resistance of Mg implants could be significantly improved by a simple method using non-toxic materials. The increase in corrosion resistance implies the possibility of using less bulky fixation plates, thus enhancing the potential of using Mg as degradable implants.
URI: http://hdl.handle.net/10397/22091
ISSN: 0925-8388
DOI: 10.1016/j.jallcom.2008.01.102
Appears in Collections:Journal/Magazine Article

Access
View full-text via PolyU eLinks SFX Query
Show full item record

SCOPUSTM   
Citations

129
Last Week
0
Last month
1
Citations as of Dec 1, 2017

WEB OF SCIENCETM
Citations

115
Last Week
0
Last month
1
Citations as of Dec 10, 2017

Page view(s)

61
Last Week
1
Last month
Checked on Dec 11, 2017

Google ScholarTM

Check

Altmetric



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.