Title: Application of high-resolution radiochromic film dosimetry in verifying a small-field stereotactic radiosurgery plan
Authors: Lee, KY
Fung, KKL 
Kwok, CS
Keywords: Film dosimetry
High-resolution film dosimetry
Microdensitometer
Radiochromic film
Issue Date: 2006
Source: Applied radiation and isotopes, 2006, v. 64, no. 8, p. 934-939 How to cite?
Journal: Applied Radiation and Isotopes 
Abstract: A high-resolution radiochromic film dosimetry (Hr-RCFD) method has been applied to verify a small-field stereotactic radiosurgery (SRS) plan. This was done by exposing a RCF in a Perspex head phantom undergoing the same treatment plan as the patient. The dose distribution obtained by the Hr-RCFD was verified against that calculated by the stereotactic treatment planning system and the result was satisfactory. The Hr-RCFD method has been found to be an accurate and practical tool in verifying small-field SRS plans.