Please use this identifier to cite or link to this item: http://hdl.handle.net/10397/18401
Title: Spiral gradient endpoint susceptibility testing : a fresh look at a neglected technique
Authors: Pong, R
Boost, MV
O'Donoghue, MM 
Appelbaum, PC
Keywords: MICs
Haemophilus
Spiral plater
Reproducibility
Evaluation
Issue Date: 2010
Source: Journal of antimicrobial chemotherapy, 2010, v. 65, no. 9, p. 1959-1963 How to cite?
Journal: Journal of antimicrobial chemotherapy 
Abstract: Objectives: Increasing antibiotic resistance and interest in matching antibiotic therapy with pharmacokinetic/ pharmacodynamic characteristics of isolates has led to increasing demands for determination of MICs. This can lead to increased costs for the laboratory. The spiral gradient endpoint (SGE) technique, a low-cost method of MIC determination, was developed some years ago. Although the technique showed good correlation with reference methods, it was not widely employed, mainly due to the introduction of alternative methods. We have revisited this technique and evaluated it for the determination of MICs for fastidious organisms.
Methods: The SGE method was first optimized for fastidious organisms using Haemophilus influenzae. Intrabatch and inter-batch reproducibility was determined for H. influenzae, Streptococcus pneumoniae, Moraxella catarrhalis and Neisseria gonorrhoeae. The method was then evaluated by comparison of MICs for clinical isolates of these organisms determined by SGE with those determined with the reference method.
Results: Optimization of the technique resulted in a method with excellent reproducibility for all organisms tested [SD 0.10–0.337; coefficient of variation (CV) 8.59% –18.66%]. These SDs/CVs were lower than those of the reference methods (0.27 –2.34; 31.0%–63.8%). There was excellent correlation of the MICs with the reference methods (0.908 –0.930) and insignificant differences in numbers of strains in each resistance category, with no tendency for SGE to produce higher or lower MICs than the reference method (P.0.05).
Conclusions: SGE was shown to be reproducible and produced results that correlated well with standard techniques for fastidious organisms. The method offers a rapid, flexible, cost-effective alternative for smaller laboratories and for routine use in developing countries.
URI: http://hdl.handle.net/10397/18401
DOI: 10.1093/jac/dkq239
Appears in Collections:Journal/Magazine Article

Access
View full-text via PolyU eLinks SFX Query
Show full item record

SCOPUSTM   
Citations

5
Last Week
0
Last month
1
Citations as of Apr 25, 2017

WEB OF SCIENCETM
Citations

2
Last Week
0
Last month
0
Citations as of Apr 26, 2017

Page view(s)

19
Last Week
2
Last month
Checked on Apr 30, 2017

Google ScholarTM

Check

Altmetric



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.